Portable respiratory diagnostic solution with highly accurate V-Core flowhead technology. Easy to use and no PC required.
Designed for testing in an office or on the go, the Alpha is our all-in-one desktop spirometry solution.
Part of our range of V-Core™ spirometers, the Alpha encompasses compliance with the ATS/ERS 2019 Technical Standards, multiple testing options, quick loading integrated thermal printer and direct links to Vitalograph Device Studio to generate PDF reports. Integration with Spirotrac 6® respiratory diagnostic software enables test data to be centralised for detailed reporting and trending. Vitalograph Connect provides versatile connectivity to your electronic medical records via HL-7 and GDT, taking spirometry testing to the next level. Features a 7” high-resolution touch-screen with an intuitive user-friendly interface.
The Alpha Spirometer comes with:
The Alpha Spirometer comes with:
Test Modules | Description | Parameters | Onscreen Graphs |
---|---|---|---|
Forced Vital Capacity | The total volume of air that can be exhaled during a maximal forced expiration effort. | FVC, FEV1, FEF25-75, PEF, etc |
|
Vital Capacity | The total volume of air that can be exhaled during a normal expiration effort. | VC, IC, IVC, TV |
|
Bronchodilator Responsiveness Testing | Determines the degree of improvement of airflow in response to bronchodilator administration as measured by changes in FEV1 and FVC. | VC, FVC, FEV1, FEF25-75, PEF, etc |
|
This device comes with a Free 5-Year Warranty including unlimited technical support when registered within 30 days of purchase.
Vitalograph provides free certified training for all of its respiratory diagnostic solutions. Register here.
Click here for reimbursement codes for Vitalograph Devices in the USA.
The Alpha has an integrated V-Core™ flowhead, allowing you to focus on helping your patients achieve their best spirometry test, confident that you are capturing accurate and reproducible results.
Find out more about V-Core flowhead technology
Product code: 69002
| Vitalograph Model 6000 Alpha |
---|---|
Flow detection principle | Fleisch type pneumotachograph |
Essential performance test limits | Flow accuracy ±10% or ±10L/min of the reading (ISO 23747:2015) |
Back pressure | Less than 0.1kPa/L/second @ 14L/s, complies with ATS/ERS 2019 |
Volume detection | Flow integration sampling @ 100Hz |
Volume accuracy | +/- 2.5% ATS/ERS 2019, ISO 26782:2009 |
Voltage/Frequency | Power Supply: Input 100-240V, 50/60HZ, Output 12V, 1.5A, Battery: 7.2V, 2.2Ahr |
Accuracy when operated in temperature range conditions | Flow ±10%, Max. flow rate ±16 L/s, Min. flow rate ±0.02 L/s |
Operating temperature range | ISO 26782 limits: 17–37ºC Design limits: 10–40ºC |
Operating humidity range | 30%–75% |
Ambient pressure range | 850hPa–1060hPa |
Performance standards met by Vitalograph Alpha | ATS/ERS 2019, ISO 23747:2015 & ISO 26782:2009 |
Safety standards | EN 60601-1:2006 +A1:2013 +A2:2021 |
QA/GMP standards | EN ISO 13485, FDA 21 CFR 820, SOR/98-282 & Japan’s PMD Act |
Dimensions | 204 mm (length) x 253 mm (width) |
Weight | 1.5 kg net |
Communications | USB (Device Studio & Spirotrac 6), Ethernet/WiFi (Connect) |
EMR Connectivity | HL7, GDT via Vitalograph Connect |
Device Classification | IIa per Annex IX of the MDD |
Test parameters measured | VC Test: VC, EVC, IVC, TV, IRV, ERV, IC FVC Test: VC, FVC, FIVC, FIVC/FVC, FEV0.5, FEV0.5/FVC, FEV0.75, FEV0.75/FVC, FEV1, FEV1R, FEV1/VC, FEV1/EVC, FEV1/IVC, FEV1/FVC, FEV1/FIVC, FEV1/FEV6, FEV1/PEF, FEV3, FEV3/VC, FEV3/FVC, FEV6, PEF L/s, PEF L/min, FEF25, FEF50, FEF75, FEF0.2-1.2, FEF25-75, FEF25-75/FVC, FEF75-85, FIV1, FIV1/FVC, FIV1/FIVC, PIF L/s, PIF L/min, FIF25, FIF50,FIF75, FIF50-FEF50, FEF50-FIF50, MVVind, FMFT, FET, FRC, TV, RV, TLC, IRV, ERV, IC, Rind, ELA, tPEF, Text, EV, EV/FVC, |